Membrane characterization by the contact angle technique: II. Characterization of UF-membranes and comparison between the captive bubble and sessile drop as methods to obtain water contact angles
Several commercial UF-membranes are characterized by water contact angles. Two different methods of contact angle measurements, the sessile drop and the captive bubble methods, are compared. Differences in results between the two methods occur for the DDS GR 61 membrane; otherwise the methods give rather consistent results. The advancing and the receding contact angles are measured for the UF-memb