Utilisation of a Pd-MOS sensor for on-line monitoring of dissolved hydrogen in anaerobic digestion
The use of a hydrogen-sensitive palladium-metal oxide semiconductor (Pd-MOS) sensor in combination with a membrane for liquid-to-gas transfer for the detection of dissolved hydrogen was investigated. The system was evaluated with known concentrations of dissolved hydrogen in water. The lowest concentration detected with this set-up was 160 nM. The method was applied to monitoring of a laboratory-s